DDS 560 Dilution System

Dynamic Dilution System DDS 560 for variable-adjustable dilution of aerosols at variable air flow rates

The dynamic dilution system DDS 560 serves for customised and reproducible dilution of aerosols with variable dilution factor and for any flow rate in the range of (0.5 – 3.0) L/min.

The adjusted dilution factor is shown on a display. In the applied principle that corresponds to the standard VDI 3491-15, particles are removed from the sample according to the specified dilution ratio – the particle size distribution remains unchanged.

Optical particle counters show a maximum measurable aerosol concentration due to their operating principle. If this limit is exceeded major errors cannot be excluded, for example because of coincidence. In case of clean room testing or laminar flow box testing according to VDI 2083 measurements of high upstream concentrations are required. Conventional clean room particle counters are unable to measure at such high concentrations.

By placing the dynamic dilutor DDS 560 before a particle measuring device, e.g. the LAP 322: Laser Aerosol Particle Size Spectrometer, the aerosol concentration can be reduced by a factor selected by the user. Thus, the required upstream concentration can be diluted to a concentration which is within the measuring range of the particle counting device.

Request a quotation for Topas products

Standards

  • VDI 3491-6
  • VDI 3491-15